3-22
3.5.2 Automatic Setting of N-0 Compensation Values
N-0 compensation values are set while increasing the rotating speed of the torque detector from zero to
the rpm specified as CAPACITY. In order to minimize the influence of the inertia of the object under
measurement, a sufficiently long time (about 30 seconds) shall be taken as the time from zero to the
maximum rpm (specified as CAPACITY). Whether or not compensation values are used can be specified
by the ON/OFF setting.
Operating Procedure
(1)
Steps (1) to (4) below are the same as those in 3.5.1, "Manual Setting of N-0
Compensation Values."
Refer to 3.5.1, "Manual Setting of N-0 Compensation Values."
(5)
1
Select the ON/OFF setting of N-0 compensation.
Select 1:ON or 2:OFF using the
and
keys.
1:ON
Use N-0 compensation.
0:OFF
Does not use N-0 compensation.
To set N-0 compensation values automatically, select 2:OFF, obtain data, then select 1:ON.
2
Select MODE.
Select 0:AUTO using the
and
keys and then press the [ENTER] key.
When you press the [ENTER] key, automatic sampling of N-0 compensation values starts. Increase the
rpm of the torque detector slowly from zero to CAPACITY (rated rpm) of REVO (rpm).
When the rpm reaches CAPACITY of REVO, data is established by pressing the [EXIT] key.
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...