4-8
• GATE INT
Setting range: 0.1 to 10s
This setting indicates the gate time when GATE selection is 0:INT, which is the data update timing in
case of LCD display and BCD OUT (GATE INT).
The data is the average value with the gate time of 4ms sample data.
If this setting is less than 1s, however, only BCD OUT follows the setting and the LCD display is updated
at 1-second intervals. The update timing of the display data for MODE1-3 depends on the GATE
selection.
In case of GATE 0:INT
The 4-ms interval sample data averaged by AVE TIME is used to display the MAX, MIN, and P-P data
within the time specified by GATE INT.
In case of GATE 1:EXT
The 4-ms interval sample data averaged by AVE TIME is used to display the MAX, MIN, and P-P data
between the external trigger (TRIG IN) and the next one.
• External trigger
Generated by pressing the [TRIG] switch on the front panel or short-circuiting REMOTE function TRIG
IN pin and common.
• AVE TIME
Setting range: 0.004 to 10s
Data average time MODE is applied to the MAX, MIN, P-P, and RIPPLE data.
TS-3200A are sampling data at 4ms intervals. The MAX, MIN, P-P, and RIPPLE data are extracted from
the 4-ms interval sample data averaged by the specified time.
Note: The average time cannot be set in 0.001s steps because of the relationship with internal
sample (4ms).
After setting the time, when you press the [ENTER] key, the time is changed to a setting
which can be processed.
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...