8-5
8. Specifications
Setup Section
Signal selection
REVO1 (MP910), REVO2 (MP981), INTSIG (torque detector SIG2)
Capacity
1 to 99999 (limited by the number of teeth of the detector and ratio
setting)
Unit
r/min, r/s, Hz
Minimum measurement rpm setting
Rpm corresponding to 1Hz input frequency
≤
Setting < Capacity
Number of teeth of detector
1 to 99999 P/R
Ratio setting
±
1 to 9999/1 to 9999
Rpm offset setting
Effective only for r/min
±
1 to 9999
Measured value = Actual measurement value - Rpm offset
Response
Time constant setting
16ms/31ma/63ms/125ms/250ms/500ms/1s/2s/4s/8s/16s/32s/64s
OFF is added only when the TS-0321 option is mounted.
Measurement Unit
Measurement clock
4.195MHz
Sampling clock
Sampling time for one data 1/256Hz (about 4ms)
Display Section
Number of display digits
5 digits
Unit
r/min, r/s, Hz
Display gate time setting
1s to 10s (shared with the torque unit) or external gate
Accuracy
±
0.05% / full-scale
±
1 count (with 1s gate time)
Output Section
Analog output format
Voltage output
0 to
±
10V/full scale
(2-phase signal input: CCW for minus output)
Scale
Attenuation setting of the full scale is possible from 0.1V to 10V in 0.1V
steps.
Response
16ms to 64s depending on the time constant setting
OFF is provided only when the TS-0321 option is mounted.
Accuracy (1s average value)
±
0.1% / full scale
Temperature drift
±
0.01% / full scale /
°
C
Appropriate load resistance
10k
Ω
or higher
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...