5.3.3 Timing Chart
Since the data/polarity output is counted for each GATE setting, data may not be updated even if the hold
signal is canceled and then held again within the specified GATE time.
Since the pulse width of the DAV command output is as short as about 5
µ
s, it cannot be read by the
general-purpose I/O port of the sequencer. When reading the BCD data and polarity, be sure to perform
hold input using the sequencer.
●
When hold input is not used (Repeating the following pattern continuously)
Since the BCD data is updated when the DAV command is LO, read the data at the rising edge.
TS-3200A
Sequencer, etc.
Data output, polarity output
DAV command output
Hold input
Data input
DAV command input
Hold output
Data output, polarity output
DAV command output
Busy input
Print operation
Printing
About 100 to 150
μ
s
Hold state
1 s, 10 s, or TRIG IN input
20
μ
s MAX