5-12
5.2.4 Recommended Interfaces
For the interface of the REMOTE function TRIG IN used in the PEAK and EXT modes, refer
to 4.1, "Using REMOTE Function."
■
CMP UUT (Contact Output) Interface
The following interface circuit is recommended.
Output format
Withstand voltage
Load power supply
ON resistance
OFF resistance
Dry contact
30VDC or less
100mA or less
10
Ω
or less
500k
Ω
or higher
Pin No.
1
2
3
4
5
6
CMP-1a
CMP-2a
CMP-1c, CMP-2c
CMP-3a
CMP-4a
CMP-3c, CMP-4c
F u n c t i o n
CH1
CH2
CH1, CH2
CH3
CH4
CH3, CH4
CMP-1a, 2a
3a, 4a
COM
V+
Pulled up with 10k
Ω
.
CMOS or TTL
TS-3200A output
External device