4-25
4.12.1 Converting Unit System from kgf to N
When using a torque detector with the kgf unit system purchased before, it is necessary to convert the
value of FACTOR and CAPACITY. However, unit system conversion from kgf to N is possible only for
detectors with a CAPACITY inscription of 10 gfcm or more.
Conversion Procedure
■
Converting Numeric Value and Unit of CAPACITY (1 kgfm = 9.807Nm)
1
Convert the numeric value inscribed as CAPACITY into kgfm.
10gfcm = 0.0001kgfm
1kgfcm = 0.01kgfm
2
Multiply the value obtained in
1
by 10. (Make setting assuming that 0.1kgfm equals
1Nm.)
3
When the value obtained in
2
is:
Less than 1
→
Multiply the value by 1000, set the result as CAPACITY, then set the unit to mNm.
Equal to or greater than 1 and less than 1000
→
Set the value as CAPACITY and the unit to Nm.
Equal to or greater than 1000
→
Divide the value by 1000, set the result as CAPACITY, then set the unit to kNm.
■
Converting FACTOR Value
1
Calculate new FACTOR by multiplying FACTOR inscribed on the detector by 0.9807.
New FACTOR = Old FACTOR x 0.9807
2
Set the result as FACTOR.
4. Other Operations
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...