8-3
8. Specifications
Output Section
Analog output format
Voltage output 0
±
10V/F.S
Scale
Attenuation setting of the full scale is possible from 0.1V to 10V in 0.1V
steps.
Response
16ms to 64s depending on the time constant setting
OFF is provided only when the TS-0321 option is mounted.
Accuracy
When combined with the detector (with 1s average value)
N-0 compensation used:
±
0.2% / full-scale
N-0 compensation not used:
±
0.5% / full-scale
Temperature drift (TS)
±
0.01% / full scale /
°
C
Appropriate load resistance
10k
Ω
or higher