8-4
8.2
Rotational Signal Measurement Unit
Input Section
■
Sine Wave of REVO1 MP910, etc.
Amplifier
AC amplification
Input impedance
10k
Ω
or higher
Input frequency range
10Hz to 100kHz
Input signal amplitude range
0.2Vrms to 45Vrms
Connector
BNC
■
Square Wave of REVO2 MP981, etc.
Amplifier
Unbalanced DC amplification
Input impedance
10k
Ω
or higher
Input frequency range
1Hz to 200kHz (guaranteed accuracy is 10Hz or more)
Input signal amplitude range
HIGH level:
+4 to +30V
LOW level:
0.6V or less
Pulse width:
2
µ
s or more
Phase judgement
When REVO2 (Double) is selected as 90-degree phase difference signal
input
When SIG2 leads SIG1 by 90 degrees:
CW
When SIG2 lags behind SIG1 by 90 degrees:
CCW
Power supply
12VDC 100mA
Connector
R03-R6F (Tajimi Electronics)
■
INTSIGN Torque Detector SIG2