5-44
5.6.11 Check Commands
TSTn
Performs self test. (The settings differ from the TS-3600B).
n = 0: Displays the ROM TEST version.
1:RAM TEST
2:LCD TEST
3:KEY TEST
4:I/O TEST
5:BCD TEST
Outputs a service request at the end.
ICSn1n2n3[,d]
Analog output calibration
If the option in [ ] is omitted, only the status is changed but calibration is not performed.
n1 =
OFF (measurement mode)
1 to 2 Calibration channel 3-6 are ignored.
n2 =
0 Voltage (fixed)
If (current) is received, the following are ignored:
n3 =
0 ZERO
1 SPAN
d =
Reading of digital voltmeter XX.XXX[V]
ICDn1n2n3 Read the analog output calibration level
n1 =
1 to 2 Calibration channel 3-6 are ignored.
n2 =
0 Voltage (fixed)
If (current) is received, the following are ignored:
n3 =
0 ZERO
1 SPAN
Data is returned as an integer with the decimal point ignored.