8-6
8.3
POWER (Output) Measurement Unit
Input Section
Calculated from the measured torque and rpm (basic expression)
Output (W) = 2
π
/ 60 x Torque (Nm) x Rpm (r/min)
PS = 0.7355kW
Setup Section
Capacity
1 to 99999
Unit
mW, W, kW, PS
Response
Time constant setting
16ms/31ms/63ms/125ms/250ms/500ms/1s/2s/4s/8s/16s/32s/64s
OFF is added only when the TS-0321 option is mounted.
Display Section
Number of display digits
Po 5 digits
Unit
mW,W,kW,PS
Display update gate time setting
1s to 10s (shared with torque and rotating section)
Accuracy
Torque display ac Rpm display accuracy
Output Section
Analog output format
Voltage output 0 to
±
10V/F.S
Scale
Attenuation setting of the full scale is possible from 0.1V to 10V in 0.1V
steps.
Response
16ms to 64s depending on the time constant setting
OFF is provided only when the TS-0321 option is mounted.
Accuracy (1s average value)
Torque output section ac Rpm output section accuracy
Appropriate load resistance
10k
Ω
or higher