• Use the instrument with the specified voltage. The standard voltage is 100 to 240VAC (no
selection required). Use of power other than that specified may cause damage to the instru-
ment. Before turning the power ON, make sure that the power meets the specified voltage.
• Only use fuses with the rating (current, voltage, and blowout characteristics) specified for
the instrument.
Rating: 250VAC 2A time-lag fuse
Use of fuses other than that specified may cause failure. Also, be sure to turn the power OFF
and unplug the power cord for at least one minute before replacing fuses.
• Do not operate this instrument in locations where there is gas or steam. Using this instru-
ment where there is steam or combustible or explosive gas may cause an explosion.
• Do not operate this instrument in locations where corrosive gas is present. Corrosion may
cause inferior connection resulting in failure.
• Do not use the instrument in locations subject to excessive vibration, humidity, or dust.
Because this instrument is a precision instrument, using it in locations subject to excessive
vibration, humidity, or dust may cause failure.
• Avoid using the instrument in locations with high temperature because there is a risk of fire.
Using this instrument in locations having a temperature exceeding the specified operating
temperature range may cause the instrument to catch on fire.
• Do not block the heat radiation system because there is a risk of fire if heat builds up inside
the instrument. Place the instrument away from the wall in locations with the best ventila-
tion possible.
• Never dismantle or disassemble the instrument. Do not remove the casing or take apart this
instrument. Use of this instrument without its casing or while taken apart may cause damage
to equipment or electric shock. When internal adjustment, inspection, or repairs are re-
quired, contact your dealer or ONO SOKKI sales office nearby.
• Do not splash or spill water on the instrument because there is a risk of fire or electric shock
because of short-circuit or heat development. If you get water inside the instrument, unplug
the power cord immediately and call your dealer or ONO SOKKI sales office nearby as
soon as possible.
• Use the instrument within the specified temperature range (0 to +40
°
C). Using or storing the
instrument in locations with rapid temperature change may cause condensation inside it
which can cause failure.
WARNING
BEFORE USING
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...