4-2
4.1
Using REMOTE Function
Note: The total length of the cable connected to the REMOTE connector shall be 5m or less. If
malfunction is caused by noise, perform shield processing as required.
4.1.1 Input (Dry Contact Input)
• Functions
CW/CCW selection:
For selection of CW/CCW from outside
Contact input:
CW when open
CCW when closed
Logic:
CW for Hi
CCW for Lo
CLR IN:
Forcibly sets both analog and digital outputs to 0 while the contact input is
closed (logic input Lo).
TRIG IN:
Selects the display or BCD data in synchronization with the signal.
Data is changed when the input to TRIG IN changes from contact open to
close (from the Hi to Lo logic).
Input interval:
100ms to 32s
COM
+5 V
10k
Ω
100
Ω
TS-3200A input
External device
Load voltage
Load current 100mAmin
100 mA min
ON resistance
10
Ω
max
OFF resistance
500 k
Ω
min
5 V
DC
min
At the time of dry contact input or logic input
At the time of contact input
Input Lo level voltage
Input Hi level voltage
3 to 5.25 V
Input impedance
1 k
Ω
0 to 1.4 V
At the time of logic input
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...