3-4
3.2
Setting Parameters of Torque Detector
Enter parameters (specific values) of the torque detector. Enter the numeric values of FACTOR,
CAPACITY, and PULSE inscribed on the name plate of the torque detector. The unit specified by UNIT
of item CAPACITY becomes the unit for the measurement display.
However, numeric values containing a decimal point cannot be entered as items CAPACITY and
FACTOR. Change values to integers before entering.
■
Example of Name Plate
■
Initial Settings
The initial settings at the time of shipment are shown below. Change these values according to the torque
detector.
• CAPACITY
The capacity refers to the rated capacity of the detector. Enter the numeric value inscribed on the name
plate of the torque detector. The value entered here becomes the full scale at the time of analog output
setting.
Input range:
±
1 to 9999 (The polarity can be changed.)
CAPACITY
+5
UNIT
Nm
FACTOR
8000
P/R
60 [P/R]
DIGIT
(number of digits)
4