4-3
4.1.2 Output (Dry Contact Output)
• Function
TRIG OUT:
Output in synchronization with the GATE time
The time specified as GATE INT in 3.8, "Setting Measurement Display."
Display and BCD are switched between at the timing indicated by
↓
.
READY output
The contact closes in the measurement mode. When using the output of the TS-3200A as a feedback
signal for control, etc., use the READY output for safety to apply interlocking.
When the following conditions are met, the measurement mode is entered.
• Conditions
1
Not the CAL Mode.
2
The TRQ SIG display is highlighted when non-DP type is used for TYPE of DETECTOR setting.
3
CLR IN of REMOTE is open.
4
Not the TEST Mode.
0.5s
0.5s
Open
Close
5s
5s
Open
Close
1s
10s
4. Other Operations