4-26
4.13
Temperature Compensation for FACTOR of Torque Detector
With phase difference torque detectors from ONO SOKKI, the material of the torsion bar at the detecting
section has temperature characteristics.
Since temperature characteristics affect FACTOR, temperature compensation after measurement is not
required if you have performed measurement using the corrected FACTOR. Since the temperature of the
torsion bar at the time of detector calibration in ONO SOKKI is inscribed on the name plate, FACTOR is
corrected automatically by entering the inscribed temperature value and the temperature value of the
torsion bar in the operating environment.
Measure the temperature of the torsion bar by bringing a thermometer such as a thermocouple into
contact with the end of the torque shaft.
Operating Procedure
(1)
Press the [MENU] key.
(2)
Select (DETECTOR SET) and then press the [ENTER] key.
(3)
Select (TORQUE) using the
and
keys and then press the [ENTER] key.
The following screen appears.