5-8
5.2.2 Description of MODE
■
AVE (0.004 to 10.000s)
The sectional average measured value for each average time setting is compared with the
comparator setting.
When the average measured value exceeds the comparison measured data, the contact
output is made.
Example: REVO 1000 r/min
The above short-circuit and open conditions apply to output logic H. For output logic L,
the open and short-circuit conditions are reversed.
■
PEAK (0.004 to 10.000s)
Peak hold is performed by the measured data for each specified average time, then the
value is compared with the comparator setting. *
Example: REVO 1000 r/min
* When the peak hold value exceeds the setting, the COMP OUT terminal of REMOTE is short-circuited.
If the measured data is less than the setting when TRIG signal is input next time, the COMP OUT
terminal is opened.
The above short-circuit and open conditions apply to output logic H. For output logic L,
the open and short-circuit conditions are reversed.
Open
Short- circuited
Short- circuited
Open
Open
CMP OUT
1000r/min
r/min
Comparator setting
• Sectional average
measured value for each
average time setting
Peak hold value
Measured value
1000r/min
r/min
Comparator setting
TRIG IN
TRIG IN
TRIG IN
CMP OUT
Short- circuited
Open
Open
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...