4-15
4.8
Calibrating Analog Output
The TS-3200A has been calibrated before shipment.
If you calibrate the TS-3200A, be sure to use a voltmeter with an accuracy of
±
0.02% or higher.
Note: During CAL operation, the measurement ready state turns OFF.
There are two different methods for the [CAL] function.
■
EXT CALIBRATION (External Output Calibration)
When an analog output is recorded on an external recorder, if it is necessary to adjust ZERO and SPAN
positions, the external recorder (pen recorder, X-Y recorder, or the like) can be calibrated by outputting
an arbitrary voltage from the TS-3200A.
■
INT CALIBRATION (Internal Output Calibration)
By measuring the analog output with an internal voltmeter and inputting relevant numerical values to the
TS-3200A, internal compensation can be performed to enable adjustment of ZERO and SPAN.