7-2
7.1
Troubleshooting
If you think of a failure, check the following points first.
If the TS-3200A does not operate normally after check, contact your dealer or ONO SOKKI sales office
nearby.
S y m p t o m
The power supply is not
turned ON.
The system is not activated
after turning on the power.
Cannot make DISPLAY
display.
The measurement unit
display does not change.
The LCD display remains
OFF.
The torque display remains
0 .
Mismatched torque display
The torque display is not 0
without load.
Cause
The power cable is
disconnected.
The fuse has blown.
Damaged backup
m e m o r y
Failed circuit
c o m p o n e n t
Incorrect display
condition setting
Turned OFF by the
LCD ON/OFF
f u n c t i o n .
The back light is not
lit or life has expired.
No signal comes
from the torque
detector.
Incorrect DETECTOR
setting
Torque zero
operation has not
been performed.
Excessive torque was
applied resulting in
twisted detector.
There exists a static
load.
Check point
Check connection.
Check the fuse in
the fuse holder.
POWER LED is lit
and initial screen of
LCD is normal.
POWER LED is lit.
Check whether
DISPLAY setting is
OFF.
Check whether
DISPLAY setting
screen is MAX, MIN,
and P-P.
Press the LCD ON/
OFF switch.
The life of back light
is 10000 hours min.
Is the motor
supplied with the
torque detector
rotating ?
Is [TRQ SIG]
highlighted in the
measurement screen
?
42- Does DETECTOR
setting agree with
detector name plate
?
Is the torque zero
value is almost the
same as the value of
Test Results of the
detector ?
Is 0 indicated by
shaking the shaft ?
Solution
Plug power cords
securely into outlet
and inlet.
Replace fuse.
Contact sales office
or dealer.
Perform backup
memory clear
operation.
Contact sales office
or dealer.
Rotate the motor
supplied with the
torque detector. In
case of SS, the switch
changes in three
positions. Set the
switch to the
rotational direction
to be measured,
other than neutral.
Make DETECTOR
setting correctly.
Perform TRQ ZERO
operation.
Contact sales office
or dealer (to repair
the detector).
Use the detector
solely.
Remarks
4.10 "Initializing
Settings"
3.8 "Setting
Measurement
Display"
4.4 "MAX, MIN, P-P,
and RIPPLE of
Display Data"
4.11 "Turning LCD
Off"
3.3 "Selecting
Rotational Direction
(CW/CCW)"
3.2 "Setting
Parameters of Torque
Detector"
3.4 "Zero
Adjustment"
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...