3-14
3.3.3 Selecting CW/CCW Using REMOTE Function
When the REMOTE function is used, CW/CCW can be selected using an external sequence.
Change the switch of the torque detector according to the rotational direction.
Operating Procedure
(1)
Press the [MENU] key.
(2)
Select (DETECTOR SET) and then press the [ENTER] key.
(3)
Select (TORQUE) using the
and
keys and then press the [ENTER] key.
The following screen appears.
(4)
Select (ROTATION) using the
and
keys and then press the [ENTER] key.
(5)
Select 0:EXT using the
and
keys and then press the [ENTER] key.
(6)
Connect signal between pins C and D of the REMOTE connector.
CW:
Opens between C and D.
CCW:
Short-circuit between C and D.
Refer to 4.1, "Using REMOTE Function" for details on the REMOTE function.