3-35
3.9
Setting Analog Output
Any two out of torque, rotation, and POWER (output) can be output to ANALOG OUT 1 and 2 on the
rear panel. These output terminals are isolated from the torque input section and the chassis of the TS-
3200A.
The time constant (
τ
) can be set in the range from 16ms to 64s. A smaller time constant value makes the
output close to the instantaneous value, while a greater time constant value makes the output close to the
average value. With a too small time constant value, however, the torque ripple component of the detector
may appear in the output signal.
When using the MD Series and SS Series torque detector, set 63ms or more.
• Specifications
Output items:
Select from torque, rotation, and output (POWER).
Number of outputs:
2
Output connector:
ANALOG OUT 1
ANALOG OUT 2
ANALOG OUT 3 (Option: TS-0328)
Output signal:
Voltage 0 to
±
10V/F.S
Proper load resistance:
10k
Ω
or higher
Full scale:
The full-scale value of the output voltage can be changed in 0.01v steps.
For the full scale, the value specified by item CAPACITY of the
[DETECTOR SET] setting is used as the rated value.
Sample time:
4ms (1/256Hz)
Measurement is performed with a sampling time of 4ms (1/256Hz).
The output data is updated at 4ms intervals (1/256Hz) regardless of the
time constant (
τ
) setting.
3. Basic Operations
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...