3-16
3.4.1 Automatic Zero Value Setting
When 5 seconds or more have elapsed after turning on the power of the detector, perform the following
operation.
Operating Procedure
(1)
Connect the TS-3200A and the torque detector using the signal cable.
(2)
When the shaft of the torque detector is in the no-load condition, then turn on the power.
Make sure the rotational direction (CW or CCW) under measurement, then change over the upper motor
of the torque detector and the TS-3200A according to the rotational direction. When using a detector
which does not use the upper motor, such as the DP type, rotate the drive shaft (no-load condition) and
then make this setting when it is rotating at 100r/min or or higher speed.
(3)
Press the [TRQ ZERO] key.
The following screen appears.
<<< Fig. >>>
(4)
Press the [ENTER] key.
Note: Automatic zero adjustment is effective only when the above screen
1
is in the
RDY
state
(measurement ready state).
1
2