4-17
■
When INT CALIBRATION Is Displayed
(1)
Same as "
■
When EXT CALIBRATION Is Displayed."
(2)
Using the
and
keys, select (INT CALIBRATION) and then press the [ENTER] key.
The following is displayed.
Note: For the optional analog output channel (TS-0328), refer to 5.7, "Adding One Analog Out-
put Channel" in Chapter 5.
• Calibrating CH1
Connect a voltmeter to ANALOG OUT1 on the rear panel.
1
Select (CH1) using the
and
keys and then press the [ENTER] key.
2
Select 0:ZERO using the
and
keys and then press the [ENTER] key.
3
Enter the display value of the connected voltmeter using the
,
,
, and
keys and then press
the [ENTER] key.
When the voltmeter is in the - display mode, make setting like -0.005V.
4
Select (1:SPAN) using the
and
keys and then press the [ENTER] key.
5
Enter the display value of the connected voltmeter using the
,
,
, and
keys and then press
the [ENTER] key.
• Calibrating CH2
Connect a voltmeter to ANALOG OUT2 on the rear panel.
1
Select (CH2) using the
and
keys and then press the [ENTER] key.
Perform the same operations as steps
2
to
5
for CH1.
To adjust and check the scale of the recorder connected to the TS-3200A, select only 0:ZERO/1:SPAN.
4. Other Operations
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...