4-16
Operating Procedure
■
When EXT CALIBRATION Is Displayed
(1)
Press the [CAL] key.
The following screen appears.
(2)
Using the
and
keys, select (EXT CALIBRATION) and then press the [ENTER] key.
The following screen appears.
When you select 0: ZERO and then press the [ENTER] key, 0V voltage is output.
When you select 1: SPAN and then press the [ENTER] key, 10V voltage is output. Depending on the %
setting, the voltage can be output as a percentage with respect to 10V. However, if the SCALE setting is
not 10V, the setting in 3.9, "Setting Analog Output" is assumed.
Example
Setting
SCALE setup
Analog output
+100 %
10.00
10 V
+100 %
5.00
5 V
+50 %
10.00
5 V
-100 %
10.00
-10 V