8-8
8.5
Interface Unit
REMOTE
■
Items
Clear input
Contact input
Sets the measurement value forcibly to 0 when the
contact is closed.
Rotational direction selection input
Contact input
Selects the CW/CCW torque zero point.
Closes the contact when CCW is selected.
Trigger input
Contact input
Updates display and BCD at contact close when the
external gate is selected.
ORed with the front panel switch when the trigger
function is used.
Trigger output
Contact output
Turns ON/OFF in synchronization with the display
gate time.
Example: In case of 1s gate, turns ON for 0.5s and
OFF for 0.5s.
Synchronous operation of multiple TS-3200A is
possible using the trigger input and trigger output.
Ready output
Contact output
When the TS-3200A is in the torque measurement
mode, the contact closes.
■
Signal Formats
Input section
In case of non-voltage contact input
Open voltage:
5.25V or less
Short-circuit current:
1mA or less
In case of voltage input
H level:
+4 to +5.25V or less
L level:
0 to +1V
Output section
Photo-MOS relay
Load voltage:
30VDC or less
Load current:
100mA or less
ON resistance:
10ohms or less
OFF resistance:
500kohms or higher
Connector
R03-R8M
Содержание TS-3200A
Страница 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Страница 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Страница 96: ...4 30...
Страница 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Страница 150: ...6 8...
Страница 151: ...7 1 Chapter 7 Troubleshooting...
Страница 154: ...7 4...
Страница 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Страница 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Страница 170: ...8 16...