BOP-1K 091719
B-23
B.80
[SOURce:]LIST:RESolution? QUERY
LIST:RES?
Syntax:
Short Form: LIST:RES?
Long Form: LIST:RESolution?
Return Value: <value1>,<value2>,<value3>
(where <value1> and <value2> = clock period, <value3> = number of points to be used)
Description:
Identifies whether resolution is set to minimum (3933) or maximum (5900) and indicates the
clock period.
The resolution (3933 or 5900) determines the maximum number of points that will be
used for the waveform. <value1> and <value2> indicate the boundaries of acceptable dwell times:
e.g., if LIST:RES? returns either 0.000093,0.034000,3933 or 0.034000,0.000093,5900 dwell times
faster than 93
S
and slower than 34 mS will not be accepted.
B.81
[SOURce:]LIST:SAMPle:CURRent COMMAND
LIST:SAMP:CURR
Syntax:
Short Form: LIST:SAMP:CURR average,value
Long Form: LIST:SAMPle:CURRent average,value
where:
<average> = number of measurements = integer: 2, 4, 8, 16, 32, 64, 128 or 256
<value> = programmed current (Amps) for current list, or programmed voltage (Volts) for voltage list.
Description:
Samples output current.
<Value> is either programmed current in Amperes, or programmed voltage
in Volts. <Average> can only be 2, 4, 8, 16, 32, 64, 128 or 256. Entering a number other than those
permitted results in the average being set to the next lowest permitted number. The actual number of
points added to the list is <average>. The samples are taken within the window established by <aver-
age>. The timing of the first sample is not exact. The sample duration must be established by the
LIST:SET:SAMP command, otherwise a -221 ”Settings conflict” error will result
This command must be preceded by LIST:CURR, LIST:VOLT or LIST:SET:SAMPLE command, other-
wise a -221 ”Settings conflict” error will result. The entries on the dwell list must match the values on
the current or voltage list, otherwise a -226 “Not Same Length” error will result. Only eight samples
may be taken in a list. If more than eight LIST:SAMP:CURR or LIST:SAMP:VOLT commands are
included in the list, a -223 “Too Much data“ error results. Also, note that each sample uses one of the
points in the waveform table; if the total number of points in the waveform exceeds 3933, -223 “Too
Much data“ error results. (See Figures B-8 and B-9.)
B.82
[SOURce:]LIST:SAMPle:VOLTage COMMAND
LIST:SAMP:VOLT
Syntax:
Short Form: LIST:SAMP:VOLT average,value
Long Form: LIST:SAMPle:VOLTage average,value
where:
average = number of measurements = integer: 2,4,8,16,32,64,128 or 256
value = programmed current (Amps) for current list, or programmed voltage (Volts) for voltage list.
Description:
Samples output voltage.
<Value> is either programmed current in Amperes, or programmed voltage
in Volts. <Average> can only be 2, 4, 8, 16, 32, 64, 128 or 256. Entering a number other than those
permitted results in the average being set to the next lowest permitted number. The actual number of
points added to the list is <average>. The samples are taken within the window established by <aver-
age>. The timing of the first sample is not exact. The sample duration must be established by the
LIST:SET:SAMP command, otherwise a -221 ”Settings conflict” error will result
This command must be preceded by LIST:CURR, LIST:VOLT or LIST:SET:SAMPLE command, other-
wise a -221 ”Settings conflict” error will result. The entries on the dwell list must match the values on
the current or voltage list, otherwise a -226 “Not Same Length” error will result. Only eight samples
may be taken in a list. If more than eight LIST:SAMP:CURR or LIST:SAMP:VOLT commands are
included in the list, a -223 “Too Much data“ error results. Also, note that each sample uses one of the
points in the waveform table; if the total number of points in the waveform exceeds 3933, -223 “Too
Much data“ error results. (See Figures B-8 and B-9.)
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......