BOP HIPWR 091719
3-71
FIGURE 3-23. WEB INTERFACE CONFIGURE LAN PAGE
The factory default is that passwords are not needed to access password-protected areas and
none of the three passwords are established. To set a password for the first time, leave the
CURRENT field blank, enter the new password in the NEW and REPEAT fields, then click SUB-
MIT. Subsequent changes require the current password to be entered in the CURRENT field. If
a password is lost or forgotten, the factory default state (no password required) can be restored
by pressing the LAN Reset button on the rear panel of the BOP 1KW-ME (see Figure 2-1).
To change a parameter, enter the correct data and/or click the appropriate check-boxes, then
click SUBMIT to save the change or NO CHANGE to clear and reload the page without saving.
If the data is not entered within 20 seconds after the last keystroke, the data will revert to the
previous value. If the LAN CONFIGURATION password was set, you will be prompted to enter
the password before the changes are accepted.
When SUBMIT is clicked after new IP address information is entered, the front panel LAN indi-
cator blinks rapidly while the unit validates the IP address. If the address is valid, the browser
returns to the INSTRUMENT HOME page (Figure 3-22). If the address entered is already in
use, the following note appears towards the top of the page:
NOTE: IP Address is in use.
Click CONFIGURE LAN and choose another IP to correct.
An IP address that is not already in use must be entered and submitted, otherwise the browser
will lock, requiring the browser to be closed and reopened.
3.5.7.6
OPERATING THE UNIT USING WEB INTERFACE
To operate the unit, click the OPERATE INSTRUMENT box at the left. A simulated front panel
(see Figure 3-24) at the top of the screen shows 1) output voltage and current, 2) VOLT or
CURR is displayed above the MODE button to indicate whether the unit is in voltage or current
mode, respectively (** is displayed if state is unknown, web page will update within 10 seconds),
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......