1-2
BOP HIPWR 091719
1.2.1
LXI COMPATIBILITY (ME MODELS)
BOP with the ME suffix are designed to meet the LXI Version 1.4 standard, providing the follow-
ing standardized web interface capabilities for the user:
Discovery:
- MDNS (Bonjour - like the printer discovery)
- VXI-11 (National Instruments Ni Max, Agilent Io controller)
IP address support:
- DHCP
- AUTOIP
- Static IP (using Web Page)
Connections:
- SOCKET @ port 5025
- TELNET @ port 5024
Connection throughput:
- 25 milliseconds
NOTE: VXI-11 socket connection is provided for discovery. It only supports the *idn? query in
operation. Consult factory if this slow connection is required.
1.3
SPECIFICATIONS
Table 1-1 below indicates parameters that vary for different 1000 Watt BOP models; Table 1-2
lists general specifications that apply to all 1000 Watt BOP models.
TABLE 1-1. BOP 1000 WATT MODEL PARAMETERS
Model
d-c Output Range
Closed Loop Gain
Output Impedance
Voltage
(V d-c)
Current
(A d-c)
Voltage
Channel
G
V
(V/V)
Current
Channel
G
I
(A/V)
Voltage Mode
(Series R - L)
Current Mode
(Parallel R - C)
R
d-c
(mOhms)
L
(
H)
R
d-c
(Ohms)
C
(
F)
1000 WATT MODELS
BOP 6-125MG/ME
0 to ±6
0 to ±125
0.6
12.5
0.05
1.5
24
1150
BOP 10-100MG/ME
0 to ±10
0 to ±100
1.0
10.0
0.1
2.0
50
1100
BOP 10-75MG/ME
0 to ±10
0 to ±75
1.0
7.5
0.13
2.0
67
976
BOP 20-50MG/ME
0 to ±20
0 to ±50
2.0
5.0
0.40
8.3
200
371
BOP 25-40MG/ME
0 to ±25
0 to ±40
2.5
4.0
0.63
15.8
313
165
BOP 36-28MG/ME
0 to ±36
0 to ±28
3.6
2.8
1.30
25
640
103
BOP 50-20MG/ME
0 to ±50
0 to ±20
5.0
2.0
2.50
50
1250
55
BOP 72-14MG/ME
0 to ±72
0 to ±14
7.2
1.4
5.14
104
2570
33
BOP 100-10MG/ME
0 to ±100
0 to ±10
10.0
1.0
10.0
163
5000
16
NOTE: When connecting active loads, the steady-state voltage of the active load must not exceed the maximum voltage rating of the
BOP. Otherwise the overvoltage protection will shut down the power supply.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......