4-10
BOP-1K 091719
31.Send
CAL:ZERO
to prepare for calibration of S_IN_PARALLEL (slave input) control signal.
Connect a 0V ±0.1mV d-c reference to pin 3 (S_IN_PARALLEL) referenced to pin 1 (SGND)
of the PAR/SER CONTROL IN connector (A2A5J3). Set the BOP to zero output current by
sending
CAL:PAR ZERO
. Connect the DVM across the sense resistor to measure the BOP
output current and send
CAL:DATA
commands as needed (see PAR. 4.3b) to adjust the
BOP output current until the DVM reads as close to zero as possible within tolerance speci-
fied in Table 4-4 for CURRENT ZERO.
NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to
meet published specifications.
32.Connect a +10V ±0.1mV d-c reference to pin 3 (S_IN_PARALLEL) referenced to pin 1
(SGND) of the PAR/SER CONTROL IN connector (A2A5J3). Set the BOP to maximum posi-
tive output current by sending
CAL:PAR MAX
. Measure the output current using the DVM.
Send
CAL:DATA
commands as needed (see PAR. 4.3b) until the DVM reading is as speci-
fied in Table 4-3 for +FULL SCALE CURRENT.
33.Save the calibration levels by sending
CAL:SAVE 12/21/2003
. The characters at the end
are optional and represent the calibration date of the power supply. The date entered in this
manner is reported in the
*idn?
query.
4.4
CALIBRATION USING FRONT PANEL KEYPAD IN LOCAL MODE
The following paragraphs describe the screens and keys used in the calibration process.
4.4.1
UNDERSTANDING THE CALIBRATION PROCESS
The following steps describe the calibration screens, and associated keys active during the cali-
bration process. Refer to PAR. 4.4.2 for full calibration procedure.
1. From the power-up screen, press
%
to enter the General Setup menu, then highlight Cali-
bration and press
!
.
2. Enter the password (see PAR. 3.2.4.3), and press
!
, The LCD shows main calibration
screen (Figure 4-3) which includes explanations for the following function keys:
•
!
- VOLTAGE - Selects the voltage calibration of the BOP.
•
@
- CURRENT - Starts the calibration for the BOP in current mode.
•
#
- EXTRL CONTROLS - Selects input and output signal calibration used for multiple
units and the external protection.
•
$
- SAVE - Save New Calibration Setting. The information in the DATE field is saved
with the Calibration.
•
%
or
ENTER
- EXIT
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......