6. From the Waveform Settings screen, highlight End of Segments and press
@
to insert the
next segment. Use
!
to modify the following settings for the next segment. For each
parameter, use
$
to save the setting
Type
Sine
Frequency (Hz)
50
Amplitude (p-p)
8
Offset
6
Start angle
0.0
Stop angle
90.0
When all settings are complete, press
$
to save for power-up. This step create a repeating
1/4 sine wave that starts at 6V and rises to 10V over 5 ms (1/4 of 20ms period established
by 50Hz frequency). The waveform shown in Figure 3-14 has now been saved. Note that by
careful calculation of period and start/stop angle, accurate waveform simulations can be
attained.
7. Either execute the completed waveform by pressing
STANDBY
(to turn the output on), then
$
, or press
%
to return to the Saved Waveforms screen.
3.3.10.10 COPYING A WAVEFORM
After observing the precautions of PAR. 3.3.10.1, press
@
from the power-up screen to enter
the Saved Waveforms screen (Figure 3-11). Highlight one of the stored waveforms, then press
$
. Then highlight a waveform location designated
Empty
and press
$
to save or
%
to abort
the change and exit. The copied waveform may then be modified and saved with a new name if
desired (see PAR. 3.3.10.8).
3.3.11
RESET
CAUTION: Pressing RESET will cause voltage transients to appear at the output which
may damage a connected load. Switch is recessed to prevent inadvertent
activation.
Pressing
RESET
causes the power supply to execute a complete power on sequence to reset
the power supply to the power up settings (see PAR. 3.3.8).
3.3.12
OPERATOR TESTING
The Test System submenu is entered by pressing
%
from the power-up screen, then highlight
Revision/Test and press
!
. Upon power-up all tests are flagged as
Untested
. Table 3-14 lists
the tests that can be run. Highlight the choice and press
!
to run the selected test.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......