BOP-1K 091719
B-11
B.24
OUTPut:CONTrol?
QUERY
OUTP:CONT?
Syntax:
Short Form: OUTP:CONT?
Long Form: OUTPut:CONTrol]?
Return Value: {HON | LON | HOFF | LOFF | LPOFF | OFF}
Description:
Indicates operation of Remote on/off pin 2 of Trigger Port.
Returns HON, LON, HOFF, LOFF or
LPOFF, OFF to indicate how Trigger port pin 2 has been configured. (see PAR. B.23 for details).
B.25
OUTPut:MODE COMMAND
OUTP:MODE
Syntax:
Short Form: OUTP:MODE {ACTive | RESIstive | BATTery}
Long Form: OUTPut:MODE {ACTIVE | RESISTIVE | BATTERY}
Description:
Determines how the unit functions when output is turned off depending on the type of load in
use: active, resistive or battery.
See PAR. 3.3.7 for description of ACTIVE, RESISTIVE and BAT-
TERY modes. To save OUTP:MODE state for power-up, send MEM:UPD OUTP (see PAR. B.20).
B.26
OUTPut:MODE?
QUERY
OUTP:MODE?
Syntax:
Short Form: OUTP:MODE?
Long Form: OUTPut:MODE]?
Return Value: ACT | RES | BATT
Description:
Indicates whether power supply is configured for ACTIVE, RESISTIVE or BATTERY type loads.
See PAR. 3.3.7 for description of ACTIVE, RESISTIVE and BATTERY modes. Related Commands:
OUTP:MODE.
B.27
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND
CURR
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP] <exp_value>
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1
Description:
Sets programmed current level at power supply output.
This command programs output current to
a specific value; actual output current will depend on load conditions. Any data that exceeds I
MAX
causes error message -222 “data out range” to be posted to the error queue (See example, Figure B-
1.)
B.28
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] QUERY
CURR?
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]? MIN, MAX
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? MIN, MAX
Return Value:
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1
Description:
Returns either the programmed value, maximum value, or minimum value of current.
The
CURR? query returns the programmed value of current. Actual output current will depend on load con-
ditions. The CURR?MAX query returns the maximum positive current allowed for a particular model.
CURR? MIN returns minimum negative current allowed for power supply (always negative).
Related
Commands:
CURR. (See example, Figure B-1.)
B.29
[SOURce:]CURRent[:LEVel]:LIMit[:BOTH] COMMAND
CURR:LIM[:BOTH]
Syntax:
Short Form: [SOUR:]CURR[:LEV]:LIM[:BOTH] <value>
Long Form: [SOURce:]CURRent[:LEVel]:LIMit[:BOTH] <value>
where <value> is between zero and I
NOM
Description: Establishes the software limit <value> for output current, i.e., sets the maximum value of output cur-
rent that the unit will be allowed to source and sink. Requires MEM:UPD (PAR. B.20) to save for
power up. See PAR. 3.5.2.2 and Figure 3-19 for special programming considerations. (See Figure B-
5)
B.30
[SOURce:]CURRent[:LEVel]:LIMit[:BOTH]? QUERY
CURR:LIM[:BOTH]?
Syntax:
Short Form: [SOUR:]CURR[:LEV]:LIM[:BOTH]?
Long Form: [SOURce:]CURRent[:LEVel]:LIMit[:BOTH]?
Returns <value>, <value>
Description: Identifies the positive software limit, followed by the negative software limit value of output current that
the unit can source or sink.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......