3-58
BOP HIPWR 091719
3.5.3.2.1 CHANGING THE GPIB ADDRESS
The default address is 6. To change the GPIB address press
%
from the power-up screen,
then highlight Interface Settings and press
!
to enter the submenu (see Table 3-17). Highlight
GPIB Address and press
!
, then use the number keys to enter the desired address and press
$
to save. Press
$
to save for power-up or
%
to apply the changes (without saving for
power-up) and exit. Note that GPIB address is not changed when factory defaults are restored
by pressing
@
.
3.5.3.2.2 ESTABLISH COMMUNICATION LANGUAGE
Communication via GPIB port can be executed using either SCPI or CIIL protocols. Press
%
from the power-up screen, then highlight Interface Settings and press
!
to enter the submenu
(see Table 3-17). Highlight Language, press
!
, highlight SCPI or CIIL and press
$
to save.
Press
$
to save for power-up or
%
to apply the changes (without saving for power-up) and
exit.
3.5.3.2.3 CONFIGURE DEVICE CLEAR (DCL) CONTROL
The device clear (DCL) and selected device clear can be set to operate in two modes. In the
MATE mode, when the device clear is received, the output of the power supply is set to zero
volts. In the SCPI mode sending DCL or selected DCL has no effect on output voltage and cur-
rent as required by IEEE specification 488.2. The factory default value is SCPI mode.
To change the Device Clear configuration press
%
from the power-up screen, then highlight
Interface Settings and press
!
to enter the submenu (see Table 3-17). Highlight Device Clear
and press
!
, then highlight SCPI or MATE and press
$
to save. Press
$
to save for power-
up or
%
to apply the changes (without saving for power-up) and exit.
3.5.3.3
RS 232 SERIAL PORT SETUP
Paragraphs 3.5.3.3.1 through 3.5.3.3.3 below describe the setup required for the BOP to com-
municate via the RS 232C Serial bus using SCPI commands. Table 3-17 shows the menu for
Serial interface settings.
3.5.3.3.1 SELECT BAUD RATE
To change the Baud Rate configuration press
%
from the power-up screen, then highlight
Interface Settings and press
!
to enter the submenu (see Table 3-17). Highlight Baudrate and
press
!
, then highlight OFF (to disable the serial port), 9600, 19,200 or 38400 and press
$
to save. Press
$
to save for power-up or
%
to apply the changes (without saving for power-
up) and exit. Note that baud rate is not changed when factory defaults are restored by pressing
@
.
3.5.3.3.2 CONFIGURE XON/XOFF PROTOCOL
Press
%
from the power-up screen, then highlight Interface Settings and press
!
to enter the
submenu (see Table 3-17). Highlight Baudrate and press
!
, then highlight XON / XOFF and
press
!
. Highlight the desired option and press
$
to save, then press
$
to save for power-
up or
%
to apply the changes (without saving for power-up) and exit.
Enable
- Allows BOP to control when data is received (see PAR. 3.5.6.2.1 for a description of
Xon/Xoff protocol).
Disable
- Disables XON/XOFF. If both Serial XON/XOFF and Prompt functions are disabled,
Echo mode is set (see PAR. 3.5.6.2.2 for a description of echo mode).
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......