3.4
ANALOG REMOTE MODE PROGRAMMING
The BOP can be programmed remotely using analog signals applied to the remote Analog I/O
port at the rear panel (see Figure 2-1 and Table 2-10). Analog remote programming allows the
user to 1) establish voltage or current mode (PAR. 3.4.3), 2) control the main channel using the
BOP as a power amplifier (PAR. 3.4.4.1) and 3) establish the protection limits (PAR. 3.4.5). An
output signal proportional to output current is also provided for external use (PAR. 3.4.6).
Analog mode may be configured remotely using either VOLT:MODE EXT (PAR. B.111) or
CURR:MODE EXT (PAR. B.35) to enable the external reference in both voltage and current
mode. Digital programming may be restored remotely by sending either VOLT:MODE FIX (PAR.
B.111) or CURR:MODE FIX (PAR. B.35).
3.4.1
SUPPRESSING NOISE AT THE OUTPUT
When using an externally grounded signal source, particularly when the BOP is powered from a
single phase a-c input power source, undesirable noise may appear at the output. Noise sup-
pression is accomplished by connecting a 100
F ceramic capacitor, rated at 50V or 100V (e.g.,
Kepco P/N 117-1241), between GND and COM S terminals (see Figure 2-3 for local sensing or
Figure 2-4 for remote sensing).
3.4.2
REMOTE OUTPUT OFF
There are two ways to turn off the output using remote signals: (1) Remote Shutdown which
requires the unit to be turned off, then on in order to restore operation and (2) Remote Standby
which sets the output to OFF, putting the unit in STANDBY status.
Unit in STANDBY - Please Change
Output must be on for the waveform to be executed. press
STANDBY
key to set output on
then press
$
to run the waveform
Can not delete Single Segment - use
F1 to Change
Waveform must have at least one segment.
Can not delete the empty segment.
“End of Segments” was highlighted when
#
- Erase was pressed. Move highlight to valid
segment and try again.
Entry not Empty, Use F1 to change.
@
pressed while a previously saved waveform was highlighted. Use
!
to modify or
highlight Empty and press
@
for a new waveform.
Under Minimum allowed value
Frequency or period was too small.
Frequency has no sign
+/- key is not functional when modifying frequency.
Use F5 to stop waveform.
$
not valid when waveform already executing.
TABLE 3-15. ERROR MESSAGE EXPLANATIONS (CONTINUED)
ERROR MESSAGE
EXPLANATION
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......