BOP HIPWR 091719
3-57
3.5.3.1
DETERMINING WHETHER *RST COMMAND SETS THE OUTPUT OFF OR ON
When operating in remote mode, the Reset command *RST (see PAR. A.11) can be configured
to turn the output ON or OFF. Refer to PAR. 3.3.7 and Table 3-6 for a full explanation plus CAU-
TIONS and WARNINGS associated with setting the output on or off, depending on which load
type is selected.
Press
%
from the power-up screen, then highlight Interface Settings and press
!
to enter the
submenu (see Table 3-17). Highlight
*RST set Output
and press
!
, then highlight ON or
OFF and press
$
to save. Press
$
to save for power-up or
%
to apply the changes (without
saving for power-up) and exit.
3.5.3.2
GPIB PORT SETUP (MG MODELS ONLY)
The following paragraphs describe the how to configure the GPIB port. Table 3-17 shows the
menu for GPIB interface settings.
TABLE 3-17. INTERFACE SETTINGS MENU
SETTING
CHOICES
(
BOLD
= Factory Default)
FUNCTION
Interface Settings
*RST set Output
OFF
ON
OFF - Sending *RST causes output to go off (per IEEE 488.2 specifica-
tions) (see PAR. 3.5.3.3 for details).
ON - Sending *RST causes output to go on (see PAR. 3.5.3.3 for
details).
GPIB Interface Settings
Language (This setting
applies to all interfaces:
LAN, GPIB and Serial.)
SCPI
CIIL
SCPI (Standard Commands for Programmable Instruments) - see PAR.
3.6. and Appendix B.
CIIL - Provides compatibility with early Kepco power supplies and con-
trollers.
GPIB Address
0 - 30 (
06
)
Assigns address used by external GPIB controller to communicate with
BOP. GPIB address is not changed when factory defaults are restored
by pressing
@
.
Device Clear
SCPI
MATE
SCPI - sending Device Clear (DCL) or selected DCL has no effect on
output voltage and current per IEEE specification 488.2.
MATE - When device clear is received, the power supply output is set to
zero volts.
Serial Interface Settings
Baud Rate
OFF
9600
19200
38400
Selects RS 232 serial port baud rate. Baud Rate is not changed when
factory defaults are restored by pressing
@
.
XON / XOFF
ENABLE
DISABLE
ENABLE - Allows BOP to control when data is received (see PAR.
3.5.6.2.1).
DISABLE - Disables XON/XOFF. If both Serial XON/XOFF and Prompt
functions are disabled, Echo mode is the default (see PAR. 3.5.6.2.2).
Prompt
DISABLED
ENABLED
ECHO
ENAB+ECHO
ENABLE - Enables Serial Prompt (see PAR. 3.5.6.2.3).
DISABLE - Disables Prompt. If both Serial XON/XOFF and Prompt
functions are disabled, Echo mode is set (see PAR. 3.5.6.2.2).
ECHO - Enables Echo (see PAR. 3.5.6.2.2) mode.
ENAB+ECHO - Enables Serial Prompt (see PAR. 3.5.6.2.3) and Echo
(see PAR. 3.5.6.2.2) modes. Prompt is returned when the unit is ready
and any received characters are echoed back to the sender.
NOTE: To restore all interface settings defaults, press
%
from the power-up screen, highlight Interface Settings, press
!
, then
press
@
to restore defaults shown above in
BOLD
. Press
$
to save for power-up or
%
to apply the changes (with-
out saving for power-up) and exit.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......