BOP-1K 091719
B-43
B.176 SYSTem:REMote? QUERY
SYST:REM?
Syntax
:
Short Form: SYST:REM?
Long Form: SYSTem:REMote?
Return Value: {1 | 0}
Description:
Identifies whether unit is in remote mode (1) or local mode (0) during serial (RS 232) communi-
cation.
See PAR. 3.5.6.3 and Figure B-13.
FIGURE B-13. SETTING THE UNIT TO REMOTE MODE VIA SERIAL (RS 232) PORT
B.177 SYSTem:SECurity:IMMediate COMMAND
SYST:SEC:IMM
Syntax
:
Short Form: SYST:SEC:IMM
Long Form: SYSTem:SECurity:IMMediate
Description:
Initializes all NVRAM variables to factory defaults.
Empties all memory locations. See PAR. 3.5.2.2
and Figure 3-19 for special programming considerations. This command is password protected (must
be preceded by SYST:PASS:CEN (see PAR B.171).
B.178 SYSTem:SET COMMAND
SYST:SET
Syntax
:
Short Form: SYSTem:SET {CM0 | CM1 | DCL0 | DCL1 | LF0 | LF1 | RO0 | RO1}
Long Form: SYSTem:SET {CM0 | CM1 | DCL0 | DCL1 | LF0 | LF1 | RO0 | RO1}
Description:
Establishes Device Clear, Line Feed, OUTP command and Reset functions. Sending
SYST:SEC:IMM sets LF1, DC0, and RO0 (as if CM0 was sent).
Requires MEM:UPD (PAR. B.20) to
save for power up. See PAR. 3.5.2.2 and Figure 3-19 for special programming considerations.
DCL0
Device Clear functions per IEEE 488.2 Standard. (No effect on the device (power supply),
only clears internal status registers.)
DCL1
Device Clear functions identical to *RST (Output set to 0V, voltage mode and output set to
OFF except if RO1 (see below) is set.)
LF0
Line Feed not provided upon empty buffer condition
LF1
Line Feed provided if buffer is empty and a read is performed.
RO0
*RST and power up set output to OFF, 0V, 0A, voltage mode.
RO1
*RST sets output to ON, 0V, 0A, voltage mode. Subsequent VOLT and CURR commands
are immediately present at the output without sending OUTP ON (OUTP OFF and OUTP
ON function normally to turn the output off or on.)
CM0
Establishes DCL0, LF0, RO0 conditions described above. (SCPI 1997 Standard compli-
ance)
CM1
Establishes DCL1, LF1, RO1 conditions described above. (Compatible with software ver-
sions 1.2 and earlier) (See example, Figure B-14.)
B.179 SYSTem:SET? QUERY
SYST:SET?
Syntax
:
Short Form: SYST:SET?
Long Form: SYSTem:SET?
Return Value: DCL<n>,LF<n>,RO<n> where n = 0 or 1
Description:
Identifies functions established by SYST:SET command.
(See example, Figure B-14.)
*IDN?
Unit responds with KEPCO,BOP 50-20,E1234,1.66 (typical).
OUTP?
Unit responds with 0 indicating output is off
SYST:REM?
Unit responds with 0 indicating unit is in local mode.
SYST:REM 1
Puts unit in remote mode.
OUTP ON
Enables output
OUTP?
Unit responds with 1 (output on).
SYST:REM 0
Unit set to local mode.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......