BOP HIPWR 091719
1-17
1.6
FEATURES
1.6.1
DIGITAL CALIBRATION
The BOP Power Supply contains no internal adjustments. Calibration is done entirely via the
keypad or remotely via the GPIB (for MG models), LAN (for ME models) or RS 232 interface
using digital entries and a calibrated DVM, a precision d-c reference voltage source and preci-
sion shunt resistor. Calibration instructions appear on the front panel after a password is
entered; previous calibration values are saved and can be restored using the remote interface if
desired. The original factory calibration values can also be restored using the remote interface.
(Refer to Section 4.)
1.6.2
VOLTAGE/CURRENT PROTECTION
Positive and negative voltage and current protection values can be individually programmed.
Refer to PAR. 3.3.3.
1.6.3
WAVEFORMS
The BOP models have the capability to make the output follow complex waveforms. These may
be generated either externally, using an analog reference voltage (see PAR. 1.6.6), or internally,
using user-generated waveforms. The waveforms can be setup and recalled from both the front
panel display and the computer interfaces.
Up to 16 user-generated waveforms can be stored for later use. The user selects the operating
mode, positive and negative protection levels, and a count (the number of times the waveform is
to be repeated). Each waveform consists of up to 10 segments; each segment is assigned one
of six basic waveform types: square, sine, triangle, negative ramp, positive ramp and level.
Depending on the waveform type, frequency, peak-to-peak amplitude, offset, and start and stop
phase angle are then assigned by the user.
Segments at the beginning of the waveform can be programmed to be executed only once, or to
be repeated as specified by the count. A graphic representation on the LCD shows the wave-
form as segments are added and is displayed prior to execution. This feature provides unique
versatility for generating a waveform that meets user requirements. Refer to PAR. 3.3.10 for fur-
ther details.
1.6.4
SAVING AND RECALLING SETTINGS
The BOP offers 99 memory locations accessible from the front panel that can be used to store a
set of operating parameters for later use. For each location, the user can store operating mode,
output on/off, Main channel reference type and value, and protection reference type and value.
The stored settings can then be recalled to quickly program the unit to the predetermined set-
ting. Refer to PAR. 3.3.9 for further details.
1.6.5
EXTERNAL REFERENCE (ANALOG CONTROL)
An external reference, provided through the rear analog port connector (see Table 2-10), can be
used to control the output of the BOP. The external reference may be applied directly or can be
attenuated by the digital controls (see PAR. 3.4.4.2). The BOP will limit its output based on the
user configurable limits so that, in the event of a failure in the external reference, the devices
connected to the output are protected. Refer to PAR. 3.4 for further details on using external sig-
nals to control the output.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......