BOP HIPWR 091719
2-31
1. Adjust the limits to reflect the increased power capacity of the multiple unit combination as
follows:
a. From the main screen, press
%
to enter the General Setup Menu.
b. From the General Setup menu, Highlight Max/Min Settings and press
!
. Press
@
to
restore defaults. Verify that the Max/Min settings reflect the multiple unit combination. For
example, for a parallel combination the values for +Current Max, –Current Min and ±Cur-
rent Protect Max/Min now reflect the total current of the parallel combination. Voltage val-
ues and limits reflect a series combination. For 2 X 2 and 3 X 2 combinations both current
and voltage are affected. NOTE: If desired, these limits may be lowered as described for a
standalone unit (see PAR. 3.3.4).
c. Press
$
to save the settings upon power-up, then
%
to exit the General Setup menu and
return to the power-up screen
FIGURE 2-17. POWER ON DEFAULT SCREEN FOR MASTER UNIT
FIGURE 2-18. POWER ON DEFAULT SCREEN FOR SLAVE UNIT
VOLTAGE
SOURCE
VOLTAGE
0000
0.
VOLTAGE
SOURCE
0.0000