B-40
BOP-1K 091719
B.153 SYS
T
em
:
COMMunication:SERial:ECHO COMMAND
SYST:COMM:SER:ECHO
Syntax
:
Short Form: SYST:COMM:SER:ECHO {ON | OFF} or {01 | 00}
Long Form: SYSTem:COMMunication:SERial:ECHO {ON | OFF} or {01 | 00}
Description:
Enables (ON) or disables (OFF) echo mode
(see PAR. 3.5.6.2.2). Sending ON or 01 causes all sub-
sequent characters to be echoed back. Sending OFF or 00 turns off the character echo after the next
line terminator character. The *RST command has no effect on echo status. See PAR. 3.5.2.2 and Fig-
ure 3-19 for special programming considerations.
B.154 SYS
T
em
:
COMMunication:SERial:ECHO? QUERY
SYST:COMM:SER:ECHO?
Syntax
:
Short Form: SYST:COMM:SER:ECHO?
Long Form: SYSTem:COMMunication:SERial:ECHO?
Return Value: {01 | 00}
Description:
Returns 01 when echo mode is active (ON) or 00 when disabled (OFF)
(see PAR. 3.5.6.2.2).
B.155 SYS
T
em
:
COMMunication:SERial:PACE COMMAND
SYST:COMM:SER:PACE
Syntax
:
Short Form: SYST:COMM:SER:PACE {NONE | XON}
Long Form: SYSTem:COMMunication:SERial:PACE {NONE | XON}
Description:
Enables (XON) or disables (NONE) data flow control via the serial interface
(
see PAR. 3.5.6.2.1).
See PAR. 3.5.3.3 and Figure 3-19 for special programming considerations.
B.156 SYS
T
em
:
COMMunication:SERial:PACE? QUERY
SYST:COMM:SER:PACE?
Syntax
:
Short Form: SYST:COMM:SER:PACE?
Long Form: SYSTem:COMMunication:SERial:PACE?
Return Value: {01 | 00}
Description:
Returns 01 when data flow control via the serial interface is enabled (XON) or 00 when disabled
(NONE)
(see PAR. 3.5.6.2.1).
B.157SYS
T
em
:
COMMunication:SERiaL:PROMpt COMMAND
SYST:COMM:SER:PROM
Syntax
:
Short Form: SYST:COMM:SER:PROM {ON | OFF} or {0 | 1}
Long Form: SYSTem:COMMunication:SERial:PROMpt {ON | OFF} or {0 | 1}
Description:
Enables (1 or ON) or disables (0 or OFF) prompt
(see PAR. 3.5.6.2.3). Sending 1 or ON causes the
unit to return > character after the command is parsed. See PAR. 3.5.2.2 and Figure 3-19 for special
programming considerations.
B.158 SYS
T
em
:
COMMunication:SERial:PROMpt? QUERY
SYST:COMM:SER:PROM?
Syntax
:
Short Form: SYST:COMM:SER:PROM?
Long Form: SYSTem:COMMunication:SERial:PROMpt?
Return Value: {1 | 0}
Description:
Returns 1 when prompt is enabled (ON) or 0 when disabled (OFF).
(See PAR. 3.5.6.2.3.)
B.159 SYSTem:ERRor? QUERY
SYST:ERR?
Syntax:
Short Form: SYST:ERR?
Long Form: SYSTem:ERRor?
Return Value:
<int_value,string>
Description:
Posts error messages to the output queue.
Returns the next error number followed by its corre-
sponding error message string from the instrument error queue. The error queue is a FIFO (first in first
out) buffer that stores errors as they occur. As it is read, each error is removed from the queue and the
next error message is made available. When all errors have been read, the query returns 0,”No error”.
If more errors are accumulated than the queue can hold, it will overflow. The oldest errors stay in the
queue but the most recent errors are discarded. The last error in the queue will be -350,”Queue Over-
flow.” Error messages are defined in Table B-6.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...