3-30
BOP HIPWR 091719
3.3.9.2
VIEWING SAVED SETTINGS
1. Pressing
!
from the power-up screen displays the Saved Setups screen (Figure 3-10)
consisting of a list of locations 1 through 20 arranged in two columns of 10 rows. Use
T
and
R
keys to display additional columns.
2. Use the
U
and
Y
keys to highlight the desired memory location; When the desired location
is highlighted, press
!
. The details of the parameters listed in Table 3-9 are displayed.
3.3.9.3
SAVING SETTINGS AND ERASING OR MODIFYING PREVIOUSLY SAVED SETTINGS
1. Press
!
from the power-up screen to enter the Saved Setups screen (Figure 3-10).
TABLE 3-9. SAVE/RECALL MENU
PARAMETER
CHOICES
(
BOLD
= Factory Default)
FUNCTION
NOTE: DEFAULT values for empty cells are the settings of the unit at the time the save/recall menu is entered.
Mode of
Operation
Voltage
Current
External
Voltage
- Selects voltage mode. Voltage value determined by SETTING.
Current
- Selects current mode. Current value determined by SETTING.
External
- Used for customized configuration where EXT VM/CM signal at pin 2 of
I/O port controls mode; consult factory for details.
Main Channel
Reference
Internal
External
ExtRefLevel
Internal
: Any external reference applied to the Analog I/O port pin 11 is ignored
and the internal reference is established by SETTING field.
External
: The internal reference is ignored, and the external reference applied to
the Analog I/O port pin 11 is used (see PAR. 3.4.4.1).
Extreflevel
: Allows the gain of the BOP to be established by the maximum output
voltage or current introduced by the SETTING field. (see PAR. 3.4.4.2).
Main Setting
nn.nnn or
nnn.nnn
(model and parameter
dependent)
Sets the active channel setting. E.g., if Voltage mode is selected, nnn.nnn is volt-
age setting, if Current mode is selected, nn.nnn is current setting. If External mode
is selected, Main Setting is not available.
Protection
Type
Internal
External
LesserLimit
Internal
- Allows limits to be controlled by Positive and Negative values.
External
: Allows limits to be controlled by analog signals applied to the analog
port (see PAR. 3.4.5).
LesserLimit
: Allows protect limit to be automatically selected from either 1) the
external analog voltage applied to the Analog I/O port or 2) the value set in the
Positive and Negative fields. Whichever limit has a lower absolute value (closest
to zero) has effect (see PAR. 3.4.5).
Positive nn.nnn
or
nnn.nnn
This is the positive value for Voltage protect if Current mode is selected, or the
positive value for Current protect if Voltage mode is selected. If External mode is
selected, this field is not accessible.
Negative
nn.nnn or
nnn.nnn
This is the negative value for Voltage protect if Current mode is selected, or the
negative value for Current protect if Voltage mode is selected. If External mode is
selected, this field is not accessible.
Output
OFF
ON
OFF - Output disabled. The behavior of the unit when output is disabled depends
on the LOAD TYPE setting (see PAR.3.3.7).
ON - Output enabled.
Содержание BOP-ME 1KW
Страница 2: ......
Страница 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 42: ......
Страница 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 76: ......
Страница 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Страница 164: ......