Chapter 4: Basic Analyzer Operation
measurements at the time sample flow stopped. When flow is interrupted, a warning will be recorded in
the warnings/errors list.
Specifying Auto Restart Settings
If your Analyzer has a Super
iOS
, you can specify whether the Analyzer should automatically restart after various
system functions. If your Analyzer has a standard
iOS
or a Sample Inlet Block, Auto Restart is available only after
Rinse Down mode. Auto Restart does not occur if the Analyzer is configured for Grab mode prior to running a
system function By default, Auto Restart is Off.
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
Advanced Setup
button.
4. Press the
Auto Restart
button.
5. Select a function and then set it to
On
or
Off
. Available system functions will depend on what kind of
iOS
System you have. Super
iOS
options include
Rinse Down
,
Verification
,
System Suitability
, and
Linearity
.
For verification, system suitability, and linearity functions, analysis only restarts only when the protocol is
completed successfully. If any of these protocols fail, analysis does not restart.
Programming the TOC Autozero
If your application requires you to perform on-line TOC measurements, you may want to set the Analyzer to
automatically perform a TOC Autozero at a specific interval. The TOC Autozero is recommended only if the TOC
concentration in the sample stream is 50 ppb or less. The TOC Autozero function is described in more detail on
page 164.
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Press the
Menu
button, select the
Maintenance
tab, and press the
Cal/ver/Validate
button.
3. Press the
Program Autozero
button.
4. Press the
State
button and select
On
.
5. Press the
Frequency
button and use the number keypad to enter how often you want to program the
TOC Autozero function to work. The
Frequency
is measured in days; if you enter 7 the TOC Autozero will
Warning
Operating the Analyzer in On-Line mode with the Flow Sensor off can result in
erroneous measurements if there is no sample flow.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...