Chapter 2: System Description
Conductivity
Raw Conductivity Range
0.01 - 35 µS/cm
Conductivity Accuracy
±0.005 µS/cm
0.25 µS/cm
±1%
0.25 µS/cm
Conductivity Precision
<0.25% RSD
Maximum Sample Conductivity
25 µS/cm at neutral pH
Instrument
Power requirements
100-240 ±10% VAC, 100 watt, 50/60 Hz
Fuses
No user-replaceable fuses
Normal Operating Environment
Intended for indoor use only
Ambient Temperature
10 °C to 40 °C (50 °F to 104 °F)
Maximum Relative Humidity
Up to 95%, noncondensing
Maximum Altitude
2,000 m (7,546 ft)
Inputs
One isolated binary input
Outputs
Serial (RS-232) port, USB, parallel printer port,
three 4–20 mA ports, four alarms ports, one
Ethernet port
Installation/Overvoltage Category
II (protects against transients present in Category II
power)
Safety Certifications
CE, ETL listed. Conforms to UL Std. 61010-1.
Certified to CSA C22.2 No. 61010-1.
Pollution Degree
2 (normally only non-conductive pollution)
Display
Backlit Quarter-VGA touchscreen display
Size
41.9 cm height x 48.3 cm width x 27.4 cm depth
(16.5 in x 19.0 in x 10.8 in)
Weight
16.9 kg (37.2 lbs)
IP Rating
IP 45
*Note: Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
**
iOS
and Super
iOS
only. For more information, see page 96.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...