GE Analytical Instruments ©2010
39 of 226
DLM 74001-04 Rev. A
Chapter 2. System Description
System Specifications
*
Total Organic Carbon
Linear range
0.03 – 2,500 ppb TOC
Accuracy
±5% of measurement
Precision
±0.03 ppb
50 ppb
±1% > 50 ppb
Analysis Modes
Grab, On-line, On-line Averaged, On-line Timed
Analysis Time
6 minutes for continuous on-line measurements
Ozone Compatibility
50 ppb O
3
continuous
200 ppb O
3
for 2 hours daily
Sample Flow Rate (nominal)
Analysis mode: 0.25 mL/min
Fast Flush (between samples): 0.5 mL/min
External Flow Rate:
Minimum 50 mL/min
Sample Temperature
1 °C to 95 °C (34 °F to 203 °F) (withstands
short-term steam exposure)**
Sample Pressure
Up to 100 psig
Interferences
Insensitive to organic heteroatoms
Calibration Stability
Typically stable for 12 months
Display Readout
3 significant digits
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...