GE Analytical Instruments ©2010
79 of 226
DLM 74001-04 Rev. A
Chapter 4: Basic Analyzer Operation
To capture the data via the serial (RS-232) port, you will need a computer that is running a serial port
communications program, such as Hyperterminal. For information on setting up Hyperterminal to communicate
with the Analyzer, see “Using HyperTerminal” on page 179 in the Appendix.
The computer’s serial port should be set to the following:
Bits Per Second: 9600 (or a rate matching that set for the Analyzer)
Data Bits: 8
Parity: None
Stop Bits: 1
Flow Control: Hardware
Interpreting Serial Output
If configured to output 500 RL format, the Analyzer outputs the data shown in Table 4 to the Serial (RS-232) Port
for every measurement,. On configurations without conductivity measurement, Field 7 is not available.
Note:
For long distance connections (up to 1,000 feet or 305 meters), GE Analytical
Instruments recommends use of either a RS-232 current loop converter or a RS-422
converter. You can also use an Ethernet connection.
Table 4: Data Fields Output to the Serial (RS-232) Port
Field
Value
1
Mode
2
Date
3
Time
4
TOC (ppb)
5
IC (ppb)
6
TC (ppb)
7
Raw conductivity (rCond, µS/cm)
8
Temperature (°C)
9
Temperature-corrected conductivity (tCond, µS/
cm)
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...