Chapter 5: Password Protection and DataGuard
Disabling Password Protection
To disable Password protection follow these steps:
1. Log in to the Analyzer with the User ID and Password.
2. Press the
Menu
button and select the
DataGuard
tab.
3. Press the
Disable Password
button. When the confirmation window displays, press
OK
.
Password protection is immediately disabled.
The minimum User Level required to access each of the menu screens is listed in Figure 22: Menu Map — Menu
Lockout Off (Requires Calibration Passwords) on page 102.
Using DataGuard
DataGuard offers support for regulation 21 CFR Part 11, providing multiple levels of user access to the Analyzer
via up to 100 unique User IDs and maintaining an audit trail of all user operations. DataGuard must be
purchased as an upgrade from GE Analytical Instruments. Once it is enabled, it cannot be disabled.
Enabling DataGuard
DataGuard is locked until it is activated by following the steps below. When activating DataGuard, to ensure
system security we strongly recommend that you create a new administrator-level account and inactivate the
default administrator account (Steps 7 through 21 below).
1. Insert the USB flash drive containing the DataGuard activation into the Analyzer’s USB port.
1. Press the
Menu
button, select the
Maintenance
tab, and press the
Advanced
button.
2. Press the
Activate Options
button.
3. Press the
OK
button and wait for the Analyzer to detect the USB flash drive.
4. Press the
Activate
button to enable DataGuard.
When the DataGuard
Activation Code
is accepted, the Analyzer security settings are immediately put
into effect and you will be required to log in to the Analyzer with the default administrator User ID and
Password.
User ID: ADMIN
Password: GEAI
5. Press the
Login
button and enter the default administrator Login ID. Press
Enter
.
6. Enter the Password and press
Enter
.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...