GE Analytical Instruments ©2010
137 of 226
DLM 74001-04 Rev. A
Chapter 6: Calibration and System Protocols
System Suitability Worksheet
Company Name ____________________________
Date _________________________
Analyst Name ______________________________
Firmware Version ______________
Analyzer Serial Number ______________________
Standards Expiration Date _______
Standards Set Lot No. (optional) ________________
Standards Vials Lot Nos. _________
Average TOC of Reagent Water (Rw)
____________________
(
Line 1
)
Average TOC of 500 ppb Sucrose Standard (Rs)
____________________
(Line 2)
Average TOC of 500 ppb Benzoquinone Standard (Rss)
____________________
(Line 3)
Response Efficiency
____________________
(Line 4)
Response Limit
____________________
(Line 5)
Response Limit = Rs - Rw
Acceptance criteria for USP System Suitability is response efficiency between 85% and 115%
Pass
Fail
Response Efficiency = (Rss - Rw)
(Rs-Rw)
-------------------------
100
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...