GE Analytical Instruments ©2010
86 of 226
DLM 74001-04 Rev. A
Chapter 4: Basic Analyzer Operation
Figure 17: The Maintenance Tab
Displaying Consumables Status
The
Main
screen displays an overview graphic that indicates the relative life status of the Analyzer’s
consumables: UV lamp, pump heads, and resin bed. Status indicators on the
Main
screen change color as the
need for replacement nears.
To display more detailed information about consumables status, follow these steps:
1. Select the
Maintenance
tab.
2. Press the
Consumables
button.
The status of each of the consumables is displayed, with an estimate of remaining life.
If you need to install new consumables, follow the instructions in Chapter 7: Maintenance for each consumable
you install; also be sure to follow the instructions in the section called “Setting the Installation or Date for New
Consumables” on page 149.
Setting the Analyzer Clock and Time Zone
You can set the Analyzer clock by following these steps:
1. Select the
Maintenance
tab.
2. Press the
Clock
button.
3. Press each date component button, enter the appropriate value, and press the
Enter
button.
4. Press each time component button, enter the appropriate value, and press the
Enter
button.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...