GE Analytical Instruments ©2010
140 of 226
DLM 74001-04 Rev. A
Chapter 7: Maintenance
Figure 32: Interior Overview of the Analyzer
Checking and Refilling the DI Water Level
The water level in the DI water cartridge should be checked periodically and refilled as necessary. For this
procedure, you will need the water bottle (from the accessories kit) filled with DI water.
To check the level of water in the DI water cartridge:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Turn off power to the Analyzer with the main power switch.
3. Open the Analyzer front panel.
4. Locate the DI water cartridge at the center of the Analyzer (see Figure 32 on page 140).
Table 13: Replacement Schedule for Consumables
Description
Typical Operating Life
DI Water Level
Refill as needed (check when replacing other items)
UV lamp
6 months
Pump tubing (with pump heads)
12 months
DI Water Cartridge
Typically 24 Months (depends on water quality)
Inline filter
Replace as needed (depends on water quality)
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...