GE Analytical Instruments ©2010
130 of 226
DLM 74001-04 Rev. A
Chapter 6: Calibration and System Protocols
2. If the Analyzer is taking measurements, stop analysis.
3. Press the
Menu
button, select the
Maintenance
tab, press the
Cal/Ver/Validate
button, and then press
the
JP Protocol
button.
4. If you have a
Super IOS
system, the
Select Rinse
screen appears. Select one of the following rinse
options, and then press the
Nex
t button:
• No Rinse
• Before
• After
• Before and After
5. If you have selected
No Rinse
or
After
, and there are vials in the Super
iOS
, remove them now. Press
Next
to continue. Wait for the Analyzer to drain each of the Super
iOS
vial chambers (about 2 minutes).
6. If you have selected to run a rinse, the Analyzer will prompt you to do one of the following:
• Insert the Super
iOS
Rinse vial set into the Super
iOS
. Press
Next
to continue.
• Or, if using individual rinse vials, press
Next
for additional instructions. Insert the individual vials in
the Super
iOS
, and then press
No Set
to proceed.
7. Do one of the following:
• If you have a purchased vial set, insert the
JP Protocol
standards cartridge into the Super
iOS
System with the label facing away from the Analyzer and press
Next
. Proceed to Step 11.
• Insert the individual vials into the Super iOS system in the following order:
Port 1 = Leave empty
Port 2 = Reagent Water (Rw) Blank Standard
Port 3 = SDBS Standard
Port 4 = KCL Standard (Optional, only if the pharmacopeia is set to include JP TOC
and
Conductivity)
Press the
Next
button, and then press the
No Set
button. Proceed to step 9.
8. If you have a standard
iOS
system:
• Slide the
iOS
system door open, and then wait for the water to drain (about 30 seconds).
• Open the vials bag and avoid touching the septa to protect against introducing foreign particles.
• Insert the Reagent Water (RW) standard into the
iOS
System, and then press
Next
.
• When prompted by the Analyzer, remove the RW standard and insert the SDBS standard into the
iOS
System, and press
Next
.
NOTE:
Open the cartridge or vials bag and avoid touching the septa to protect
against introducing foreign particles.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...