Chapter 3: Installation
•
GMT Difference
— Allows you to enter the offset from Greenwich Mean Time. Use the number pad
and the
+/-
button to specify the offset. For example, for Eastern Standard Time you should enter
-
5.00
.
Naming the Analyzer Location (Optional)
You can assign a name to the Analyzer that displays with printed and exported data. This feature is particularly
useful if you have multiple Analyzers at your facility and want to easily distinguish data collected from each unit.
To assign a name, follow these steps:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
Advanced
Setup
button.
4. Press the
Location
button. Use the keypad to specify a name for the Analyzer and press
Enter
. Press
the
Number
button and then the
Alpha
button to toggle between numbers and letters.
5. Press the
Back
button twice to return to the
Maintenance
tab.
Setting the Analyzer Mode
Before starting normal Analyzer operation, confirm that the mode settings match your configuration’s needs.
1. Select the
Setup
tab. Confirm that the correct mode is selected.
• In
On-Line
mode, the Analyzer measures TOC continuously and returns measurement results every
6 minutes.
• In
On-Line Timed
mode, the Analyzer returns a single TOC measurement at specific times of the
day.
• In
On-Line Averaged
mode, the Analyzer, returns the averaged value of TOC over a specific time
interval.
• In
Grab
mode, the Analyzer draws samples from a vial in the Analyzer’s
iOS
or Super
iOS
System
(from vial port 1). Grab mode is not available for configurations with the Sample Inlet Block.
For more information on mode selection, see “Setting the Analyzer Mode” on page 62.
2. Press the
Menu
button to exit and save your settings.
Setting Up the Data History
Before taking measurements, review the
Archive Data
setting to ensure that data is collected in the best way
for your environment. When
Archive Data
is set to
On
, the Analyzer will not overwrite the oldest data and
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...