Chapter 8. Troubleshooting
Overview
This chapter provides the starting point for troubleshooting basic issues with your Analyzer. If you need
additional assistance, in the U.S. contact GE Analytical Instruments Technical Support at 800.255.6964. In other
parts of the world, contact your local GE Analytical instruments representative.
Step 1: Review Warnings and Errors
The best first step in troubleshooting erratic readings, poor reproducibility (>3%), or other operational problems
is to review and print the Warnings/Errors list.
To display and print the Warnings/Errors list, follow these steps:
1. Select the
Maintenance
tab.
2. Press the
Warnings/Errors
button.
3. Use the scroll buttons to browse the warnings and errors.
Note:
For Analyzer firmware with the DataGuard feature, a Maintenance-level (or higher)
User Level is required for performing most diagnostic tests; a QA-level (or higher)
password is required for performing the 4-20mA test. Check with your system
administrator for an appropriate password. For more information, see “Using
DataGuard” on page 105.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...