Chapter 6: Calibration and System Protocols
6. Do one of the following:
• If you have a purchased vial set, insert the
Suitability
standards cartridge into the Super
iOS
System
with the label facing away from the Analyzer and press
Next
.
• If you have purchased individual vials rather than a vial set in a cartridge, confirm the value shown
on the Conductivity Standard label. Next, make sure the vials are inserted into the
iOS
System vial
ports in the following order:
Port 1 = RW Reagent Water Control
Port 2 = Rs Standard Solution
Port 3 = Rss System Suitability Solution
Port 4 = Rinse Water
Press the
Next
button, and then press the
No Set
button. Go to step 10.
7. If you have a standard
iOS
System:
• Open the door to the
iOS
System and wait 30 seconds for water to drain.
• Insert the Reagent Water (Rw) into the
iOS
System and press
Next
.
• When prompted, remove the Reagent Water (Rw) from the
iOS
System, insert the 500 ppb Sucrose
Standard (Rs), and press
Next
.
• When prompted, remove the 500 ppb Sucrose Standard (Rs) from the
iOS
System, insert the 500 ppb
Benzoquinone Standard (Rss), and press
Next
.
• When prompted, remove the 500 ppb Benzoquinone Standard (Rss) from the
iOS
System, insert the
Rinse Water standard, and press
Next
.
• Press the
button if you want to print the Calibration results.
8. After the last standard has been analyzed, the System Suitability summary screen is displayed (see
Figure 31). The Analyzer indicates if the system suitability passed or failed based on the following
criteria:
Acceptance criteria for USP System Suitability is response efficiency between 85% and 115%.
Response efficiency is calculated as follows:
Response Limit is calculated as follows:
Response Limit = Rs - Rw
9. Click the
Apply
button if you want the new data to be saved as the Analyzer’s response limit.
If you accept the Response Limit, this new value determines when to trigger the Analyzer’s Response
Limit alarm, when activated (see “Setting Alarm Values” on page 83). Press the
Yes
button to confirm
that you want to accept the new value.
10. Press the
Exit
.
RE = (Rss - Rw)
(Rs-Rw)
-------------------------
100
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...